AllenNeuralDynamics / AllenNeuralDynamics/brainwasher

Minor Cross-Contamination from slight "Backwash"

未关闭
#7 1 条评论 0 个 reaction 已指派 1 人 已被 @Poofjunior 认领 在 GitHub 查看
主要语言
Python
星标
0
派生
1
PR 合并指标
30 天内没有已合并 PR

描述

Running a protocol produces small droplets as "backwash" into some lines.

![pressure_eq_bug](https://github.com/user-attachments/assets/9e38e032-137d-4984-8c23-bb2b722788c9)

This issue has to do with switching between chemicals on the rotary shear valve while having a difference in pressure between the chemical source line and the pump. It could also be an artifact due to the dead volume of gas (i.e: compressible) between the syringe and the rotary shear valve.

There are a few approaches before we can consider this "fixed."
* If we can prevent the primed line from moving altogether when we switch ports on the rotary shear valve, that's ideal.
* If we can remove the residual liquid such that there is no cross contamination but the actual primed line moves slightly, that's ok too.

贡献指南

这个仓库没有索引到贡献指南

评估

这个 Issue 还没有评估数据。

把新 issue 发到你的邮箱

精选适合新手参与的 GitHub issue 摘要。