Lakeshore218: Show disabled sensors as disabled

Aperta
#3,772 1 commento 0 reazioni 0 assegnatari Vedi su GitHub
Lingua principale
Nessun dato sulla lingua
Stelle
6
Fork
2
Merge medio
16h 40m
PR unite (30g)
2

Descrizione

As an instrument scientist, I want to specify which Lakeshore 218 sensors are disabled and not see any readings from them so that I don't get confusing readings which are over 400K from the Lakeshore 218 from sensors which are not connected.

Currently, the Lakeshore 218 reads values from disabled channels. These channels then show high temperatures as the PV values even though it says disabled on the front of the Lakeshore 218 box. These high-temperature readings are confusing for instrument scientists/users.

This ticket arose from helping TOSCA instrument scientists after migrating TOSCA to IBEX.

### Acceptance Criterion

- [ ] I can disable a sensor on the IOC level and not get a reading from the Lakeshore 218.
- [ ] The block value of a disabled sensor is sensible (e.g. 0)
- [ ] The PVs associated with a channel which is disabled do not go into an alarmed state.

Guida per i contributori

Nessuna guida per i contributori indicizzata per questo repository

Direzione di ricerca

Parti dalla configurazione dei sensori a livello IOC di Lakeshore 218 e traccia il modo in cui i canali disabilitati producono letture e allarmi PV. Il lavoro è completato quando i sensori disabilitati non producono alcuna lettura, il valore del loro blocco è sensato, ad esempio 0, e i relativi PV di canale non entrano in uno stato di allarme.

Scritto dal modello di indicizzazione a partire dal testo della issue.

Valutazione

Ambito
embedded-iot
Tipo di issue
Bug
Difficoltà
4/5
Tempo stimato
3-5 giorni
Stato di attività
Ferma
Chiarezza
Abbastanza chiara
Idoneità per principianti
45/100

Ricevi le nuove issue nella tua casella

Un breve riepilogo di issue GitHub adatte ai principianti.