Lakeshore218: Show disabled sensors as disabled
- Lingua principale
- Nessun dato sulla lingua
- Stelle
- 6
- Fork
- 2
- Merge medio
- 16h 40m
- PR unite (30g)
- 2
Descrizione
As an instrument scientist, I want to specify which Lakeshore 218 sensors are disabled and not see any readings from them so that I don't get confusing readings which are over 400K from the Lakeshore 218 from sensors which are not connected.
Currently, the Lakeshore 218 reads values from disabled channels. These channels then show high temperatures as the PV values even though it says disabled on the front of the Lakeshore 218 box. These high-temperature readings are confusing for instrument scientists/users.
This ticket arose from helping TOSCA instrument scientists after migrating TOSCA to IBEX.
### Acceptance Criterion
- [ ] I can disable a sensor on the IOC level and not get a reading from the Lakeshore 218.
- [ ] The block value of a disabled sensor is sensible (e.g. 0)
- [ ] The PVs associated with a channel which is disabled do not go into an alarmed state.
Guida per i contributori
Nessuna guida per i contributori indicizzata per questo repository
Direzione di ricerca
Parti dalla configurazione dei sensori a livello IOC di Lakeshore 218 e traccia il modo in cui i canali disabilitati producono letture e allarmi PV. Il lavoro è completato quando i sensori disabilitati non producono alcuna lettura, il valore del loro blocco è sensato, ad esempio 0, e i relativi PV di canale non entrano in uno stato di allarme.
Scritto dal modello di indicizzazione a partire dal testo della issue.
Valutazione
- Ambito
- embedded-iot
- Tipo di issue
- Bug
- Difficoltà
- 4/5
- Tempo stimato
- 3-5 giorni
- Stato di attività
- Ferma
- Chiarezza
- Abbastanza chiara
- Idoneità per principianti
- 45/100