As a MuSR scientist, I would like to be able to enable the IOC to check whether a recondense is required and if it is the IOC recondenses automatically. I would also like to be able to define the values used for autorecondense condition checks.
## Acceptance criteria:
- When recondense required conditions are met then a recondense is triggered
- When recondense required conditions are not met then a recondense is not triggered
## Recondense conditions:
All of the first set of conditions must be met:
- ITC503 is using the control channel chosen for the He3PotLo
- He3PotLo Setpoint < Use He3 below operational parameter defined by user/scientist
- Autorecondense enabled by user/scientist
One of the second set of conditions must be met:
- Sorb temperature check is enabled (by user/scientist) and Sorb temperature > condense threshold for the sorb temperature defined by the user/scientist
- He3Pot temperature check is enabled (by user/scientist) and He3Pot temperature > condense threshold for the He3 Pot temperature defined by the user/scientist
- Temperature error check is enabled (by user/scientist) and the difference between He3Pot temperature setpoint and He3Pot actual temperature > condense threshold for temperature error defined by the user/scientist