## SHP1-205241

Calibration coefficient: 9.1394 (from manufacturer 2020-06-02)

WRR from 2021 IPC: 1.010177

Baltic Pyrheliometer Comparison [report](https://library.wmo.int/viewer/69637/download?file=IOM_146_en.pdf&type=pdf&navigator=1)
Applied WRR: 1.010177 (from 2021 IPC)
Calculated adjustment: 1.000164 (validates the 2021 WRR)

## SHP1-185163

"SHP1-185163 had a calibration in 2018, sensitivity = 8.518 uV/W/m2." From Kipp & Zonen
SHP1-185163: 8.52 (datasheet value from 2018) used for IPC measurements
SHP1-185163: 8.61577 (2025-02-13 Kipp & Zonen calibration)

Contributor guide

No contributing guide indexed for this repository

Research direction

The issue lists calibration values for SHP1-205241 and SHP1-185163, but names no file, test, entry point, or requested format. First locate where station calibration history is stored and ask the maintainers which records and representation should be added; done means the intended history is recorded accurately and can be verified.